Error Messages
PR 5220 Instrument Manual
EN-156 Sartorius
13.4 Error messages of the Calibration
The maximum capacity is too high (the calculated input voltage for the
specified maximum capacity exceeds 36 mV).
Subsequent changing of the maximum capacity is possible; if you
decrease the capacity, a message is displayed if the new maximum
capacity is lower than the test load ([Calibrated at]).
This message displays, if the selected resolution is to low, e.g. 5 kg.
The selected resolution is so high that less than 0.8 internal counts per
scale interval (d) or 0.5 µV/e for legal-for-trade acc. to OIML/NSC are
available.
The maximum capacity is not an integer multiple of the scale interval.
Weight units do not match, e.g. subsequent change of [Max] from kg to
lb.