Error Messages
PR 5220 Instrument Manual 
EN-156  Sartorius 
13.4  Error messages of the Calibration 
The maximum capacity is too high (the calculated input voltage for the 
specified maximum capacity exceeds 36 mV). 
 
Subsequent changing of the maximum capacity is possible; if you 
decrease the capacity, a message is displayed if the new maximum 
capacity is lower than the test load ([Calibrated at]). 
This message displays, if the selected resolution is to low, e.g. 5 kg. 
 
The selected resolution is so high that less than 0.8 internal counts per 
scale interval (d) or 0.5 µV/e for legal-for-trade acc. to OIML/NSC are 
available. 
 
The maximum capacity is not an integer multiple of the scale interval. 
 
Weight units do not match, e.g. subsequent change of [Max] from kg to 
lb.