4.4.13 Parameter Input
Open the menu via -[Weighingpoint]-[Calib]-[Param].
Weighingpoint/WP A/Calibration
Measuretime
160 ms
Digital filter
bessel
External supply
8 - 12 VDC
Fcut
3.00 Hz
This line is shown only if the filter is on.
Test mode
absolute
W & M
none
Standstill time
0.50 s
Standstill range
1.00 d
[Measuretime]
Select the measuring time; possible values: 5, 10, 20, 40, 80, 160, 320, 640, 960, 1280, 1600 ms.
In 'legal-for-trade' mode select 1 s.
[Digital filter]
The digital filter can be switched on only with the measuring time set to 160 ms.
Select the filter characteristic [off], [bessel], [aperiod.], [butterw.], [tscheby.].
[Fcut]
Enter the cut-off frequency for the filter (0.180 Hz); the setting is dependent on the measuring time.
[External supply]
With external load cell supply (e.g. 7.5 V DC via 1626/60 with MX8 = closed), [below 8 VDC] can be selected to
adapt the sense voltage monitoring to the lower supply voltage.
[Test mode]
With [Absolute], the test value is calculated when the test is called. With [relative], the deviation from the
initially stored test value is displayed; see Chapter 4.4.11.
[W & M]
Setting for 'legal-for-trade' mode; select [none], [OIML], [NTEP] (for USA) or [NSC] (for Australia),
see Chapter 4.4.13.1.
[Standstill time]
Settings [Standstill range] and [Standstill time] are required for determining the mechanical stability of the
scale. Input in seconds; permissible range 0.01 bis 2 sec. (max. measuring time x 32). If 0 is set, stability is not
checked. The stability time must not be less than the measuring time and not greater than 32 times the
measuring time.
[Standstill range]
The scale is stable as long as any changes in the weight value are within this range; permissible range:
0.01d to 10.00d. In 'legal-for-trade' mode select 1 d.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Sartorius PR 5220 Series and is the answer not in the manual?

Sartorius PR 5220 Series Specifications

General IconGeneral
BrandSartorius
ModelPR 5220 Series
CategoryTransmitter
LanguageEnglish

Related product manuals