EasyManuals Logo

Sartorius PR 5220 Series User Manual

Sartorius PR 5220 Series
172 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #92 background imageLoading...
Page #92 background image
Commissioning
PR 5220 Instrument Manual
EN-92 Sartorius
4.6.10 Finishing/Saving the Calibration
Press to complete the calibration.
Unless all data were determined during calibration using [New] (e.g. dead load
not set / entered), the following prompt is shown:
After replying [Yes] to confirm and pressing , another prompt is displayed:
Press [Save] to save changes in calibration data. If you press [Undo], changes
are not saved and the display returns to the selection menu for the weighing
points.
The Confirmation is displayed with:
Saving calibration
After quitting the menu, the following message is displayed:
Exit calibration
After finishing the calibration, set the CAL switch to the closed position; see also Chapter 4.1.1.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Sartorius PR 5220 Series and is the answer not in the manual?

Sartorius PR 5220 Series Specifications

General IconGeneral
BrandSartorius
ModelPR 5220 Series
CategoryTransmitter
LanguageEnglish

Related product manuals