PR 5220 Instrument Manual Extended Functions
Sartorius EN-151
10.2 Updating a new Software with ‚FlashIt’
Updating the software is only possible with fixed IP address!
Always flash/load the BIOS into the device first, and then the firmware&application.
Updating the Software in a Network/via Point-to-Point Connection with a Fixed IP Address
- Device and notebook/PC are connected to the network/each other.
- The automatic address assignment ‘DHCP’ is deactivated in the device and in the
notebook/PC, see Chapter 16.2/16.3.
- The ‘FlashIt!32’ program (in a directory on the enclosed CD-ROM) is installed on the
notebook/PC.
- The ‘FlashIt!32’ program is started.
Device presettings in the setup menu
Setup/Network parameter
1.
Select -[Network parameter].
HW address 00:90:6C:6B:6A:5E
Hostname PR5230-6A6B5E
Use DHCP
2.
Deactivate ‚Use DHCP’.
IP Address 172.24.20.101
3.
Enter the corresponding IP address*.
Subnetmask 255.255.0.0
4.
Enter the corresponding subnet mask.
Default gateway 0.0.0.0
Remote access
VNC-Client 255.255.255.255
5.
Leave with .
* Using a point-to-point connection the notebook/PC has to be set to a fixed IP address too. This IP address
must be in the same number range, which was configured by the Subnetmask.
Example:
PR 5220: IP address 172.24.22.1
Notebook/PC: IP address 172.24.22.2
Device and notebook/PC have the Subnetmask 255.255.255.0.