Commissioning
PR 5220 Instrument Manual 
EN-82  Sartorius 
4.5.10  xBPI Calibration with Built-in Weight 
Prerequisites: 
-  The xBPI protocol has been selected (see Chapter 4.5.1).  
-  The weighing point ‚xBPI-scale’ has been selected (see Chapter 4.5.4). 
-  The platform configuration has been executed (see Chapter 4.5.3).  
-  The setting in the following menu has been done: [Weighingpoint A-xBPI-Scale]-[Setup]:  
[Configuration]-[Confirming adjust.]: [manual] (see Chapter 4.5.5). 
-  The communication between instrument and platform is active. 
Weighingpoint/xBPI-Scale Setup 
Calibration   
Open with  . 
Select group of specification 
 
 
Weighingpoint/xBPI-Scale Setup 
Adjust with default weight 
  Adjust with intern weight   
Select with   and enter with  . 
The procedure is shown e.g. with the following message: 
Weighingpoint/xBPI-Scale Setup 
The calibration progress is displayed. 
 
 
 
Weighingpoint/xBPI-Scale Setup 
Select [Accept]. 
 
 
The data are saved and the instrument generates a corresponding message: 
Weighingpoint/xBPI-Scale Setup 
Leave with  .