PR 5220 Instrument Manual Commissioning
Sartorius EN-85
4.6.5 Adjustment Sequence
During calibration, no data is changed in the digital load cells. The calibration data and parameters are saved
in the instrument. The unique serial numbers of the connected load cells are monitored.
For the adjustment is the following order specified:
Search for load cells, see Chapter 4.6.6.
Assign load cells, see Chapter 4.6.7.
Calibrate ‘New’: Max with weight unit, scale interval, dead load, calibration weight, see Chapter 4.6.8.
Note:
For further information about calibrating weighing points, see Chapter 4.4.
4.6.6 Search for Load Cells
Weighingpoint
Weighingpoint A
Pendeo Load Cells
Press [Assign].
Setup
Config
Param
Assign
Weighingpoint/WP A
Type
Pendeo process
No of LC
1
LC 1
0
Press [Search].
Search
View
Calib
?
Search for connected load cells.
The actual settings are reset to
default.
Cancel
Continue
Press [Cancel] to accept and display the existing values.
Press [Continue] to start a new search process.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Sartorius PR 5220 Series and is the answer not in the manual?

Sartorius PR 5220 Series Specifications

General IconGeneral
BrandSartorius
ModelPR 5220 Series
CategoryTransmitter
LanguageEnglish

Related product manuals