PR 5220 Instrument Manual Commissioning
Sartorius EN-85
4.6.5 Adjustment Sequence
During calibration, no data is changed in the digital load cells. The calibration data and parameters are saved
in the instrument. The unique serial numbers of the connected load cells are monitored.
For the adjustment is the following order specified:
• Search for load cells, see Chapter 4.6.6.
• Assign load cells, see Chapter 4.6.7.
• Calibrate ‘New’: Max with weight unit, scale interval, dead load, calibration weight, see Chapter 4.6.8.
For further information about calibrating weighing points, see Chapter 4.4.
4.6.6 Search for Load Cells
Weighingpoint A
Pendeo Load Cells
Press [Assign].
Setup
Config
Param
Press [Search].
Search for connected load cells.
The actual settings are reset to
default.
Press [Cancel] to accept and display the existing values.
Press [Continue] to start a new search process.