Commissioning
PR 5220 Instrument Manual
EN-60 Sartorius
4.4.5 Determining the Dead Load
Note:
If the liniarization is active (see Chapter 4.4.10):
After selection of the line ‚Dead load at’ the following tip is displayed:
Can not be changed here
while linearization is active
Weighingpoint/WP A/Calibration
Max
3000 d
3000 kg
Scale interval
3000 d
1 kg
Deadload at
0.00 kg
0.000000
Max at
3000.00 kg
1.000000
Not calibrated
Sensitivity
833.33
4.000000
by load
by mV/V
CalcTest
To use the empty scale/hopper as dead load (normal case):
discharge the scale/hopper
press [by load]
After
or confirmation of the change is displayed:
Set deadload
If the mV/V value of the dead load was calculated, or if it is known from the previous calibration, the value can
be overwritten by pressing [by mV/V].
Weighingpoint/WP A/Calibration
Max
3000 d
3000 kg
Scale interval
3000 d
1 kg
Deadload at
0.00 kg
0.000000
Max at
3000.00 kg
1.000000
Calibrated at
3000.00 kg
1.000000
Sensitivity
833.33
4.000000
by load
by mV/V
CalcTest

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Sartorius PR 5220 Series and is the answer not in the manual?

Sartorius PR 5220 Series Specifications

General IconGeneral
BrandSartorius
ModelPR 5220 Series
CategoryTransmitter
LanguageEnglish

Related product manuals