PR 5220 Instrument Manual  Commissioning 
Sartorius  EN-61 
Possible error messages 
The dead load entered in mV/V plus maximum capacity in mV/V is higher 
than 3 mV/V (= 36 mV). 
 
This message displays, if the scale is not stable. 
Remedial action 
•  Check the mechanical function of the scale. 
•  Adapt the filter setting; reduce the resolution, if necessary. 
•  Adapt the stability conditions.
 
 
This message displays, if the Measurement signal is negative when determining 
the dead load with [by load]. 
Cause 
Load cells connected with wrong polarity or defective. 
 
This message displays, if dead load entered in mV/V is higher than 5 mV/V.