Specifications
PR 5220 Instrument Manual
EN-160 Sartorius
14.2 General Data
The following characteristics are valid after a warm-up time of at least 60 minutes (reference temperature: 23°C).
14.2.1 Power Supply
Supply voltage 24 V DC ±20 %
Power consumption
standard 6 Watts
with fieldbus option 8 Watts
14.3 Effect of Ambient Conditions
14.3.1 Environmental Conditions
Temperature range
Reference temperature 23 °C
Ambient temp. for operation 10,000 e -10... +40 °C
‘not legal for trade’ -10... +50 °C
Power-on temperature 0... +50 °C
Storage/transport -40... +70 °C
Humidity <95 %, no condensation, (acc. to IEC 68-2)
Protection type to DIN 40050 IP30
14.3.2 Electromagnetic Compatibility (EMC)
All data in compliance with EN 61326 industrial area
Housing Radio frequency electromagnetic fields (80…1000 MHz) EN 61000-4-3
10 V/m
Radio frequency electromagnetic fields (1.4…2.0 GHz) EN 61000-4-3 3 V/m
Radio frequency electromagnetic fields (2.0…2.7 GHz) EN 61000-4-3 1 V/m
Electrostatic discharge (ESD) EN 61000-4-2 6/8 kV
Signal and Fast transients (burst) EN 61000-4-4 1 kV
control lines Peak voltages (surge) 1.2/50 µs EN 61000-4-5 1 kV
Conducted disturbances by radio frequency (0.15…80 MHz) EN 61000-4-6 10 V
Mains input Fast transients (burst) EN 61000-4-4 2 kV
Peak voltages (surge) 1.2/50 µs EN 61000-4-5 1/2 kV
Conducted disturbances by radio frequency (0.15…80 MHz) EN 61000-4-6 10 V
14.3.3 RF Interference Suppression
Electromagnetic emission In acc. with EN 61326, limit value class A, industrial area