PR 5220 Instrument Manual Error Messages
Sartorius EN-155
13.2 Weight Error Status
Error 1
Arithmetic
Error 7
(negative)
Error 2
Overload
Error 3
(>36 mV)
Error 6
Sense monit.
Error 9
Communicat.
flashing 1Hz
flashing 1Hz
Altern.
flash. 1Hz
Altern.
flash. 1Hz
flashing 1Hz
flashing 1Hz
Altern.
flash. 1Hz
flashing 1Hz
flashing 1Hz
flashing 1Hz
flashing 1Hz
Altern.
flash. 1Hz
Altern.
flash. 1Hz
13.3 Error Messages with xBPI Scales
VNC Text Cause
No values from scale No Communication with xBPI scale.
No weight data This message is shown, when the weighing point is switched over.
Press
to assign the new weighing point to the device.
Scale not ready
This message is shown
- during the warm-up phase.
- when the device is in the automatic taring mode.
- if the device has been switched on with the scale loaded.
Switch the device off and on again.
Wrong configuration The number of load cells does not correspond to the load cell configuration
settings.
Wrong serial number ‘Serial number’ does not correspond to the number set in the device.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Sartorius PR 5220 Series and is the answer not in the manual?

Sartorius PR 5220 Series Specifications

General IconGeneral
BrandSartorius
ModelPR 5220 Series
CategoryTransmitter
LanguageEnglish

Related product manuals