EasyManuals Logo

Sartorius PR 5220 Series User Manual

Sartorius PR 5220 Series
172 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #68 background imageLoading...
Page #68 background image
Commissioning
PR 5220 Instrument Manual
EN-68 Sartorius
Weighingpoint/WP A/Calibration
Tare timeout
2.5
s
Zeroset range
50.00 d
Zerotrack range
0.25 d
Zerotrack step
0.25 d
Zerotrack time
0.0 s
Overload
9 d
[Tare timeout]
Enter a timeout value between 0.1 and 25 s for a taring/zero set command that cannot be executed (e.g., if
scale mechanically unstable, filter settings faulty, resolution too high, stability condition too narrow).
[Zeroset range]
Determine a ±range around the zero point determined by the dead load during calibration; within this range:
- the displayed gross weight can be set to zero by pressing the zero-setting key (or by a corresponding
external command), and
- automatic zero tracking is active.
Available range: 0.00 d to 10000.00d
In 'legal-for-trade' mode a value 2 % of Max. must be entered. Example: 60 d for 3000e, class III.
[Zerotrack range]
Range within which automatic zero tracking compensates deviations; 0.25 to 10000.00d.
In 'legal-for-trade' mode a value of 0.25 d has to be entered.
[Zerotrack step]
Step for automatic tracking; 0.25 to 10d. In 'legal-for-trade' mode a value of 0.25 d has to be entered.
[Zerotrack time]
Enter a time interval for automatic zero tracking within 0.0 (tracking switched off) and 25 s.
In 'legal-for-trade' mode a value of 0.25 d has to be entered.
[Overload]
Weighing range above the maximum capacity (Max), without error message. Available range: 0 to 9999900 d.
In 'legal-for-trade' mode max. 9 d=e has to be entered.
Weighingpoint/WP A/Calibration
Minimum weight
50
d
Range mode
Single range
[Minimum weight]
Minimum weight at which a print command can still be executed. Range is 0 to 9999900 d.
In 'legal-for-trade' mode min. 20 d has to be entered.
[Range mode]
For scale range selection, see Chapter 4.4.13.2.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Sartorius PR 5220 Series and is the answer not in the manual?

Sartorius PR 5220 Series Specifications

General IconGeneral
BrandSartorius
ModelPR 5220 Series
CategoryTransmitter
LanguageEnglish

Related product manuals