PR 5220 Instrument Manual  Error Messages 
Sartorius  EN-157 
 
The dead load entered in mV/V plus maximum capacity in mV/V is higher 
than 3 mV/V (= 36 mV). 
 
Remedial action 
•  Check the mechanical function of the scale. 
•  Adapt the filter setting; reduce the resolution. 
•  Adapt the stability conditions. 
Measurement signal is negative (load cells connected with wrong polarity 
or defective) when determining the dead load with [by load].  
Cause 
Load cell connected with wrong polarity, or defective. 
The dead load entered in mV/V is higher than 5 mV/V. 
 
Remedial action 
•  Check the mechanical function of the scale 
•  Adapt the filter setting; reduce the resolution. 
•  Adapt the stability conditions. 
 
 
The weight on the scale is less than the dead load after input of the 
weight value. 
 
 
The maximum capacity is not an integer multiple of the scale interval.