PR 5220 Instrument Manual Commissioning
Sartorius EN-89
Corner correction will be reset.
Press [No] for no corner correction reset.
Press [Yes] to reset the corner correction and calibrating continues.
Weighingpoint/WP A/Calibration
Number of vessel feet (in the case, 4).
Local gravity
[Number of vessel feet]
Enter number of vessel feet.
[Local gravity]
Enter local gravity (in this case: Hamburg 9.81379 m/s
2
); see e.g.
http://www.ptb.de/cartoweb3/SISproject.php.
[Max]
The load cell capacity is suggested as Max (E
max
* number of load cells).
The maximum capacity (Max) determines the maximum measured weight without dead load. Normally, the
selected Max must be smaller than the load cell capacity (nominal load * number of load cells) – dead load, in
order to prevent overloading the load cells.
Adapt the value for Max. When entering Max, the required number of digits behind the decimal point must be
taken into account. Press
to switch units.
[Scale interval]
Select the scale interval (1, 2, 5, 10, 20 or 50). The number of intervals is calculated and fed back
automatically.