Commissioning
PR 5220 Instrument Manual
EN-94 Sartorius
Weighingpoint/WP A/Calibration
Zerotrack range
0.25
d
Zerotrack step
0.25 d
Zerotrack time
1.0 s
Overload
9 d
Min
20 d
Range mode
Single range
[Zerotrack range]
Range within which the automatic zerotrack is balanced.
Setting range: 0.25…10000.00 d
For use in Legal-for-Trade mode a value of 0.25 d must be entered.
[Zerotrack step]
If a weight change exceeds the adjusted value, automatic tracking does not function any more.
Setting range for automatic tracking increments: 0.25... 10 d
For use in Legal-for-Trade mode a value of 0.25 d must be entered.
[Zerotrack time]
Time interval for automatic zero tracking.
Setting range: 0.0…25 s
At 0.0 s the tracking is switched off. For use in Legal-for-Trade mode a value of 1 s must be entered.
[Overload]
Weighing range above the maximum capacity (Max) without error message.
Setting range: 0…999900 d. For use in Legal-for-Trade mode a value of max. 9 d = e must be set.
[Min]
Minimum weight at which a print command can be triggered.
Setting range: 0…999900 d. For use in Legal-for-Trade mode a value of at least 20 d must be set.
[Range mode]
For scale range selection, see Chapter 4.4.13.2.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Sartorius PR 5220 Series and is the answer not in the manual?

Sartorius PR 5220 Series Specifications

General IconGeneral
BrandSartorius
ModelPR 5220 Series
CategoryTransmitter
LanguageEnglish

Related product manuals