PR 5220 Instrument Manual Commissioning
Sartorius EN-53
4.3.8.3 Weighingpoint ‚ Pendeo Load Cells
- Search
Search load cells
- Search for a new network and reset the load cell data to the factory
settings.
- Or replace a single faulty load cell with the existing data.
- View
Serial number and current weigh of connected load cells are displayed.
- Info
Data of selected load cell are displayed.
- Assign
Load cells (serial number) are assigned to the mounting place.
- Calib
Calibration xBPI load cells
- New
Corner correction will be reset
Yes, No
- Number of vessel feet
Enter number of vessel feet
- Local gravity
Enter local gravity (Default: Hamburg 9,81379 m/s
2
)
- Max
Max. capacity: 0.000010 ...<3000>... 9999998 <kg>, t, lb, g, mg, oz
- Sccale interval
<1>, 2, 5, 10, 20, 50
- Dead load
Empty scale
- CAL weight
Calibration weight: 0.000010 ... 9999998 <kg>, t, lb, g, mg, oz
- Corner correction
O.k., when realized
- Modify
For minor modifications/ subsequent dead load setting only, otherwise,
use [New].
siehe ‚New‘
- Param
- Ambient conditions
Very stable, stable, unstable, very unstable
- W & M
<none>, OIML, NSC, NTEP
- Standstill time
0.01 s…<0.50 s> ... 2.0 s
(The range depends on the measurement time.)
- Standstill range
0.00 d ... <1.00 d> ... 10.00 d
(The range depends on the measurement time.)
- Tare timeout
0.1 s ... <2.5 s> ... 25 s, timeout when there is no standstill
- Zeroset range
0.00 d ... <50.00 d> ... 10000.00 d
- Zerotrack range
0.00 d ... <0.25 d> ... 10000.00 d
- Zerotrack step
0.00 d ... <0.25 d> ... 10.00 d
- Zerotrack time
<0.0 s> ... 25 s
- Overload
0 d ... <9 d> ... 999900 d
- Min
0 d ... <50 d> ... 999900 d, min. weight
- Range mode
<Single range>, Multiple range, Multi-interval
- Range limit 1
0…max. weight, transition from small to medium scale interval; only
when multiple range or multi-interval has been selected.
- Range limit 2
0…max. weight, transition from medium to large scale interval; only
when multiple range or multi-interval has been selected.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Sartorius PR 5220 Series and is the answer not in the manual?

Sartorius PR 5220 Series Specifications

General IconGeneral
BrandSartorius
ModelPR 5220 Series
CategoryTransmitter
LanguageEnglish

Related product manuals