Name Description Remark Ref.
EXT.SV TYPE Not used in CL model.* Initial value: 0 -
EXT.SV SERIAL Not used in CL model. Initial value: - -
EXT.SV
CHANNEL
Not used in CL model.* Initial value: 0 -
PRESSURE UNIT
Sets the unit of the displayed
pressure.
Initial value: 0
Set value:
0: MPa 1: kgf/cm
2
2: bar 3: psi
P.42
INNER
VOLUME
Input the tube volume.
Initial value: 0
Set value: 0 to 9999.9
P.43
* Do not change the factory default value.
n [UTILITY] setting group
Name Description Remark Ref.
KEY CLOSE Disables the key input. - P.44
BRIGHTNESS
Sets the brightness of display
screen.
Initial value: 4
Set value: 1 to 4
P.44
BEEP MODE Sets the operation of buzzer.
Initial value: 0
Set value:
0: ON 1: Alarm 2: OFF
P.44
VOLUME Sets the buzzer volume.
Initial value: 2
Set value: 1 to 3
P.44
DISP OFF TIME
The operation panel turns off after
the time set here elapses.
Initial value: 1
Set value: 0 to 10 min
P.44
DIRECT KEY
MODE
Specifies whether to display the
confirmation screen when the
direct key is operated.
Initial value: 0
Set value:
0: Confirm 1: Direct
P.45
n [EDIT] group
Use this group to create a time program. For details, see "3.7 Creating Time Program"
P.69.
n [FILE] operation group
This function is not used because events are not used in the CL model.
Name Description Remark
FILE NUMBER Selects a program file No.
Initial value: 0
Set value: 0 to 9
30
LC-40D X3 CL LC-40D XR CL
3 Operation

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Shimadzu LC-40D XR CL and is the answer not in the manual?

Shimadzu LC-40D XR CL Specifications

General IconGeneral
BrandShimadzu
ModelLC-40D XR CL
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals