The table shows the messages displayed in the second row of a maintenance log.
Message Description Ref.
INITIALIZE PARAM The parameters were initialized. P.62
PARAM LOCK ON The parameter lock function was turned ON.
P.62
PARAM LOCK OFF The parameter lock function was turned OFF.
PARAM LOGIN
When the parameter lock function was ON, a user
entered the password and moved on to the next
screen.
CHANGE PASSWORD The password was changed. P.63
USE PARTS MGMT The PARTS MGMT TOOL was set to "Use".
P.63
NOT USE PARTS MGMT The PARTS MGMT TOOL was set to "NotUse".
n Displaying the error logERROR LOG
This displays ten records of previous errors
with the dates of the errors.
After press [], press [] several times to
display the error log records in order.
The figure shows an example that
"a maximum pressure error occurred on
April 1, 2019".
For the messages displayed in [ERROR
LOG], refer to the System Guide.
System Guide "5.4 Error Log List"
n Displaying the operation time of the degassing unitDGU OP TIME
The screen displays the operation time of
the degassing unit connected to the
instrument.
When the operation time is input, count
of the operation time of the degassing
unit starts from the input time. The input set value is stored in the maintenance log.
If [999999 h] is set, the display shows "******" and the operation time will not be
counted.
To activate this function, supply power from the instrument to the degassing
unit and make sure that the pressure signal cable is connected correctly. If
more than one pump is used, be sure to connect the power cable and pressure
signal cable to the same pump. For how to connect, refer to the System Guide.
The shutdown function of CBM-40 CL or LC-Workstation can stop the degassing
unit. For details, refer to the instruction manual of the system controller or
LC-Workstation.
Approximate time to replace the vacuum pump of the degassing unit is 3 years
or 8000 hours of operation time. Contact your Shimadzu representative if you
need to replace the vacuum pump.
52
LC-40D X3 CL LC-40D XR CL
3 Operation

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Shimadzu LC-40D XR CL and is the answer not in the manual?

Shimadzu LC-40D XR CL Specifications

General IconGeneral
BrandShimadzu
ModelLC-40D XR CL
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals