Operating the measuring device
58 © SICK AG • Subject to change without notice • 8024484/2019-06-06
“SpecFu” menu
(cont.)
Parameters Description
AvgSpd Selectlterdepthforthemeasuredspeedvalues.
Options
Medium
Slow
Fast
Factory setting
Medium
MsrDir Deningthemeasuringdirection:
If the measuring direction is positive “+”, the distance output
value is equal to the internal distance value measured by the
measuring module. If the measuring direction is negative “-”,
the internal distance value is multiplied by a factor of “–1”.
Thechangeindirectionaectsboththeoutputvalueforthe
distance and the output value for the speed.
Options
+
Factory setting
+
ErrRej Select time for error suppression. During this time, the old
measured value is output. If there is still no valid measured
value once the time selected for the parameter “ErrRej” has
elapsed the value “0” is output.
Options
200 ms: The error/warning is reported when the error is
indicated for longer than 200 ms.
50 ms: The error/warning is reported when the error is indi-
cated for longer than 50 ms.
O:Theerror/warningisreportedimmediatelyandwithout
delay.
Factory setting
200 ms
Heat Requirement for display
Only for DL100-xxHxxxxx variants with heating
In this menu you can set the temperature at which the heating
turnson.Thehysteresisisxedat2K.
Adjustment range
–10 … +40 °C
Factory setting
–10 °C
Table 24: “SpecFu” menu