70 FLOWSIC100 Flare-XT · Operating Instructions · 8023761/V 1-0/2020-10 · © SICK Engineering GmbH
Installation FLSE100-XT
9 The self-cutting ring remains in the fixed position on the duct probe.
10 Then tighten the pipe screw fitting again hand-tight.
WARNING: Leakage hazard
Repeated use may damage the gasket of the pipe screw fitting.
Before reuse, i.e. whenever the pipe screw fitting is to be tightened again,
check the gasket of the pipe screw fitting:
Replace the gasket if it has visible deformations, indentations or damage. In
this case contact SICK Service.
Otherwise there is a risk of leakage.