Annex
FLOWSIC500 · Operating Instructions · 8025733/1GMJ/V4-2/2022-07 · © SICK Engineering GmbH 161
Subject to change without notice
Fig. 81 Control diagram 9215965 (page 2)
SICK Engineering GmbH
Bergener Ring 27
01458 Ottendorf-Okrilla
Verteildatum:
Page
Ersatz für:
Änderung
gez.
Tag
Name
Ind.
Tag Name
Maßstab:
Werkstoff
Ersetzt durch:
Gepr.
Ursprung:
MKO
2of7
Optionally Exia
Temperature Sensor
Optionally Exia
pressure sensor
Ultrasonic Transducers
Temperature range
and pressure range
see Marking plate
active passive
Terminal/
Connector*)
Function Uo
Voc
(V)
Io
Isc
(mA)
Po
(mW)
Co
Ca
(μF)
Lo
La
(mH)
Ui
Vmax
(V)
Ii
Imax
(mA)
Pi
(mW)
Ci
(μF]
Li
(mH)
BAT1 +/-
Terminal /
connector
Battery pack 1 or
external power
supply
-- -- -- -- -- 20 667 772 n.s. n.s
BAT2 +/-
connector
Battery pack 2 resp.
backup barrery
-- -- -- -- -- 20 6 67 320 n.s n.s
DO0 +/-
terminal
Digita l output 0
optical isolated
-- -- -- -- -- 20 -- 1100 n.s. n.s
DO1 +/-
terminal
Digital output 1 8.2 0.83 1.7 7.6 100 20 -- 1100 n.s n.s
DO2 +/-
terminal
Digita l output 2
optical isolated
-- -- -- -- -- 20 -- 1100 n.s n.s
DO3 +/-
terminal
Digita l output 3
optical isolated
-- -- -- -- -- 20 -- 1100 n.s n.s
RS485
-/-/+/+/A/B
RS485 Data
interface (option)
optical isolated
-- -- -- -- -- 20 -- 1100 1.35 0.03
P1/T1 sensor RS 485 include d
power supply
for p & T sensors
8.2 396 716 6.4 0.2 -- -- -- -- --
P2/T2 sensor SPI included
power supply
for p & T sensors
8.2 407 739 7.6 0.2 -- -- -- --. --
*) connectors for internal connection onl y n.s.: denotes negligible small
Entity paramaters for external connections in the Exi terminal compartment
[Ex ia] Associated Equipment
Only for interconnection with an Exi equipment or an [Exi] associated equipment
Entity parameters of interconnected equipments must be complied as follows:
Interconnection with interal p & T sensors approved
Vo < Vi, Io < Ii, Po< Pi, Co > Ci + Ccable, Lo > Li + Lcable or
Voc < Vmax, Isc < Imax, Ca > Ci + Ccable, La > Li + Lcable
Class I, Division 1, Groups C and D, Temp. Code T4
Ex ia IIB T4 Ga
ClassI,Zone0AExiaIIBT4Ga
.
-25°C < Tamb < 60°C
,
for extended range see Marking plate
.
In the US install in accordance with the NEC (NFPA70, Article 504)
and ANSI/ISA-RP12.06.01
In Canada install in accordance with CEC part 1
Exia Intrinsically Safe; Securite Intrinseque
2014-07-16
Wire size for all terminals: 0,14..0, 5 mm (AWG 24 ... 20)
2
4mm
(AWG 12)
2
EXT. POWER
BAT1
n.c.
DO3
DO0
+-
P1
P2
T1
T2
SENSORS
DISPLAY
LOCK
OFF ON
5..12V
NAMUR
4,5..16V
1.5 mm²
(AWG 16)
:
R:100
S
2..16V
2..16V
DO2
2..16V
DO1
+-
BAT2
+-
+- +- +-
9215965
kochami
draft
01
2014-08-27
gepr.
kochami
2014-08-27
RS485 (option, see user manual)
Control drawing FLOWSIC500 isolated I/O
kochami
Ed.6
02
WARNING: EXPLOSION HAZARD
Substitution of components may impair Intrinsic safety
AVERTISSEMENT: La substituti on
de composants peut compromettre la securite intrinseque.
RISQUE D` EXPLOSION -
2015-09-24
kochami
ZY57
03
2018-05-02