8024638/AE00/V1-0/2019-09| SICKSERVICE MANUAL | MCS200HW
Subject to change without notice
5 REPAIRS
Fig. 140: Sender/receiver unit, outer plate
14 Loosen the fastening screws of sheet metal covers 1 and 2 with a Phillips screwdriver
(4
pieces).
Fig. 141: Sender/receiver unit, sheet metal covers 1 and 2
15 Remove sheet metal cover 3.
1 Loosen the fastening screws with a Phillips screwdriver (3 pieces).
2 Remove the fastening screws with tweezers.
3 Pull sheet metal cover 3 off to the rear.
4 Pull the cable of the heating spirals out of the recess of sheet metal cover 3.
Fig. 142: Sender/receiver unit, sheet metal cover 3
16 Remove sheet metal cover 4.
1 Loosen the fastening screws with a Phillips screwdriver (2 pieces).
2 Take off sheet metal cover 4.
1Plug
1 Fastening screw, sheet metal cover 1
2 Fastening screw, sheet metal cover 2 (on the side)
1Fastening screw
2 Sheet metal cover 3
SIC_18_MCS200-151.tif