14 Device care
The following device care tasks are necessary for the scanner to function
properly:
Axis calibration
3D calibration (optional)
Cleaning the scanner
TIP
smart optics scanners do not require maintenance.
14.1
Axis calibration
The calibration of the axes of the scanner ensures the correct alignment of the
axes and thus accurate measurement results.
Schedule
Calibration is required after:
The first installation of aural Scan
An aural Scan upgrade
Transporting the scanner to a different place
Four weeks since the last calibration (with automatic reminder).
Calibrate the axes daily if there are fluctuations in temperature and/or
humidity.
Procedure
Restart aural Scan or finish the current scan.
Select Axis calibration in the menu.
You can optionally start from the info message (displayed after 30 days):