EasyManua.ls Logo

Spectra-Physics Tsunami - Page 84

Spectra-Physics Tsunami
238 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Tsunami Mode-Locked Ti-sapphire Laser
6-6
Figure 6-5: Pump Mirrors P
1
and P
2
11. Align cavity focus mirror M
2
.
Adjust M
2
vertically and horizontally to direct the small amount of
pump light reflected from M
2
through the center of M
1
.
12. Adjust the focus of M
2
and M
3
.
Use an IR viewer and white card for the following steps.
a. Place the white card in front of
M
4
.
b. Increase the pump laser to maximum power.
c. Adjust M
3
vertically and horizontally to center the image on M
4
.
There will be one or two reflected images visible on the card: one
from
M
3
and perhaps one from M
1
. If only one image is visible,
adjust
M
1
until two are present. The stationary image is from M
3
.
Note: they need not over-lap at this time.
d. Remove the white card.
e. Temporarily remove the
M
4
mirror to allow the fluorescence to
shine on the output bezel, then tape a white card in the florescence
image on the bezel. An IR viewer may be necessary to see the
image when using the long or
x-long optics set.
Vert. Horiz.
P
1
Pump
Steering Mirror
Horiz. Vert.
P
2
Pump
Focusing Mirror
P
2
Setscrew
On bottom,
opposite this hole.
P
2
Translation
Screw
Caution!
Perform Steps e, f, h, i, and j only if you changed the focus of P
2
, M
2
or
M
3
during this alignment session, otherwise skip to the next section.

Table of Contents