EasyManua.ls Logo

Svantek SV 977D - System Check; Calibration by Sensitivity (Acoustic Signal)

Svantek SV 977D
380 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
36
SV 977D User Manual
3.3.1 System Check
Note: It is advised to perform the system check of the instrument each time before the measurements
begin. If system check gives negative result, then it is necessary to perform calibration.
3.3.2 Calibration by Sensitivity (Acoustic signal)
Calibration on the base of the microphone’s
published sensitivity information can be
performed the following way:
1. Open the Calibr. by Sensitivity position in
the Calibration list.
<ENT>
2. Set the Sensitivity of the microphone
taken from its calibration certificate using
the / key and then press <ENTER>.
The calibration factor (Cal. Factor) is
calculated every time after pressing the
/ ► key, in relation to the nominal value of
50.00 mV/Pa.
<ENT>

Table of Contents

Related product manuals