EasyManua.ls Logo

Svantek SV 977D - Calibration by Measurement (Vibration Signal)

Svantek SV 977D
380 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
38
SV 977D User Manual
6. Press <ENTER> to save the new calibration factor (Enter to Confirm), or press <ESC> to reject it. In both
cases the instrument exits the Calibration by Measurement screen.
It is recommended to repeat the calibration measurement few times. Obtained results should be almost the same
(with 0.1 dB difference). Reasons for unstable results are as follows:
the calibrator is not properly attached to the instrument,
there are external acoustic disturbances such as high noise levels close by,
the calibrator or the measurement channel (the microphone, the preamplifier or the instrument itself) are
damaged.
7. Detach the calibrator from the microphone.
Note: During the calibration measurement, any external disturbances (acoustic noise or vibrations)
should not exceed a value of 100 dB (when using a calibrator that generates a level of 114 dB).
Note: Maximum deviation of the calibration factor cannot exceed +/-1.5 dB. In other case please
contact Svantek service.
3.3.5 Calibration by Measurement (Vibration signal)

Table of Contents

Related product manuals