6-8
1234578 6
pocH-100
i
Revised June 2008
TROUBLESHOOTING
INTRODUCTION
LAN no Response
Possible cause
Computerconnectingcablefailure.
Computermainswitchnotturnedon,orcomputernotready
for communication.
LANconnectorforhostcomputerfailure.
Action to resolve the error
Checkthehostcomputer‘scable.
Checktheconnectioncabletohost.
Press"Retry"tocleartheerror,andthetransmissiontoLAN
is restarted.
Press"Cancel"tocleartheerror.AlldataforLANoutputis
deleted.
Special notes, if error re-
mains
Not ready until the error is resolved. If the retry action does
not resolve the problem, take the host transmission off-line in
order to return the instrument to ready mode.
HC ACK Timeout
HC Off-line
HC NAK Retry
Possible cause
Computerconnectioncablefailure.
Computermainswitchisnotturnedon,orcomputernot
ready for communication.
Hostcomputerserialinterfaceerror.
Action to resolve the error
Inspectthehostcomputer‘scable.
Inspecttheconnectioncabletohost.
Press"Retry"tocleartheerror,andthe transmission tothe
host computer is restarted.
Press"Cancel"tocleartheerror.Alloutputqueuedtohost
computer is deleted.
Special notes, if error re-
mains
Not ready until the error is resolved. If the retry action does
not resolve the problem, take the host transmission off-line in
order to return the instrument to ready mode.