6-7
12345 786
TROUBLESHOOTING
INTRODUCTION
pocH-100
i
Revised February 2011
Printer Error
Possible cause
Thebuilt-inthermalprinterhasanerror.
Theprintercoverisopen.
Action to resolve the error
Checkifthepaperissetcorrectlyandclosethecover(see
p a g e 4 - 11 ) .
Aftercheckingthebuilt-inthermalprinter,press"Retry"but-
ton to clear the error.
The data currently being printed will restart printing from the
beginning of that sample.
Aftercheckingthebuilt-inthermalprinter,press"Cancel"but-
ton to cancel the data being printed.
Also discard the waiting data and the error is cleared.
Special notes, if error re-
mains
Not ready until the error is resolved.
No Printer Paper
Possible cause No printer paper in the built-in thermal printer.
Action to resolve the error
Afterrellingpaperinthebuilt-inthermalprinter,press"Re-
try"tocleartheerror.
The data currently being printed will restart printing from the
beginning of that sample.
• Afterrefillingpaper inthe built-inthermalprinter,press
"Cancel"tocancelthedatabeingprinted.
Also discard the waiting data and the error is cleared.
Special notes, if error re-
mains
Not ready until the error is resolved.
6.3. ERROR MESSAGES, POSSIBLE CAUSES AND ACTIONS TO RESOLVE THE ERROR
HC Buffer Full
LAN Buffer Full
Print Buffer Full
R-Cover open
Possible cause The right side cover was opened.
Action to resolve the error
Afterclosingtherightsidecover,pressthe"Execute"button
to clear the error, and the recovery sequence is executed.
If the recovery process is completed correctly, then Auto
Rinse and background check are performed in sequence, and
the screen returns to the Main screen.
Special notes, if error re-
mains
Not ready until the error is resolved.
PLT Noise Error
RBC Noise Error
WBC Noise Error
Possible cause
• Effectsofexternalelectricnoiseinterference.Controlboard
malfunction.
• Hydrauliclineoraperturepartlyclogged.
Action to resolve the error Separate the source of electric noise from the instrument.
Special notes, if error re-
mains
Ready for next analysis, however, the following sample re-
sults may be incorrect.
Possible cause Error in the motor controller circuit.
Action to resolve the error Turn the main switch OFF, then turn it ON again.
Special notes, if error re-
mains
Not ready until the error is resolved.
PPMC Cont. Error
Possible cause
The amount of data for output or printout is too large to be
processed.
Action to resolve the error Press"Back"tocleartheerror.
Special notes, if error re-
mains
Not ready until the error is resolved.