8-4 Revision: 2.1
No.
Parameter Description Access
level
Range of values Factory
setting
21 Cycle Count
(Cycle 71)
Number of repeating program cycles
(e.g. material tests)
(optional)
3 0 – 255 0
22 CycleCtrTime
(Cycle 74)
Pause between program cycles if
“CycleCounter” > 1 (optional)
3 0 – 9999 min. 10
23 Extern Steam
(Cycle 77)
Determines whether or not the autoclave
is heated with external steam:
0 = Internal steam generator
1 = External steam
3 0.1 0
24 StartByTime
(Cycle 70)
Allows a time to be pre-programmed
(including date) at which a selected
sterilisation program should be started
3 0.1 0
25 F0 Enable
(Cycle 78)
At the end of the program, the
sterilisation effect of the heating and
cooling phases calculated using the F0
formula can be printed out. This
sterilisation time can be deducted from
the actual sterilisation time during a
program cycle:
0 = Inactive
1 = Values printed out only
2 = SterTime – heating phase
3 = SterTime – heating/cooling phase
3 0 – 3 0
26 ExhMode
(Cycle 33)
Defines the steam release after the
sterilisation phase:
0 = Quick steam release
1 = Slow steam release
2 = Ambient air cooling from 101 °C,
chamber is blown from outside
3 = Durham (controlled steam release to
101 °C, then normal cooling with or
without support air)
4 = Cooling with or without support air
5 = Cooling with controlled support air
(as the temperature sinks, the
auxiliary pressure also sinks up to
“CoolMinPress”)
6 = As 5, but with nitrogen instead of
compressed air
7 = Special glass test (ExhTime runs to
100 °C, then cooling is made up to
EndTemp)
8 = As 5, but excess pressure is reduced
via Y7 instead of Y9 (necessary for
softwater spray cooling, for example)
9 = Not used
10 = Vacuum test program
4 0 – 10 Depends
on program