Revision: 2.1 8-5
No.
Parameter Description Access
level
Range of values Factory
setting
27 SelPT100
(Cycle 17)
Determines which sensor is used as the
control sensor during the heating and
sterilisation phases:
5 = Chamber = chamber sensor (flexible
Pt100)
6 = Condense = sensor in the flow line or
exhaust filter
7 = Other chamber sensors, for example
(optionally available)
4 5 – 10 Depends
on program
28 CoolPT100
(Cycle 37)
Determines which sensor is used as the
control sensor after the sterilisation
phase:
5 = Chamber = chamber sensor (flexible
Pt100)
6 = Condense = sensor in the flow line or
exhaust filter
7 = Other chamber sensors, for example
(optionally available)
4 5 – 10 5
29 SelSensor2
(Cycle 19)
Defines whether a second sensor should
be used for control:
5 = Chamber = chamber sensor (flexible
Pt100)
6 = Condense = sensor in the flow line or
exhaust filter
7 = Other chamber sensors, for example
(optionally available)
4 5 – 10 Depends
on program
30 Temp2LimMode
(Cycle 20)
Determines the permissible positive
temperature difference between
SelSensor2 and SelSensor in the heating
and sterilisation phases:
0 = Not active
1 = SelSensor2 must not be warmer than
SelSensor by more than Temp2Val
2 = SelSensor2 must not be warmer than
the configured sterilisation
temperature by more than Temp2Val
4 0, 1.2 0
31 Temp2Val
(Cycle 21)
Maximum positive temperature
difference between SelSensor2 and the
main control sensor or the sterilisation
temperature according to
Temp2LimMode.
4 0.1 – 50.0 °C 1.0 °C