V/D
S
ERIES
P
ARAMETERS
8-6 Revision: 2.1
No.
Parameter Description Access
level
Range of values Factory
setting
32 Heat2SterMod
(Cycle 22)
Defines the transition from the heating
phase to the sterilisation phase:
0 = Transition to sterilisation phase when
the set temperature is measured on
SelSensor
1 = Transition to sterilisation phase when
the set temperature is measured on
SelSensor and (set temperature –
Sensor2Dif) is measured on
SelSensor2
2 = Transition to sterilisation phase when
the set temperature is measured on
SelSensor and the associated steam
pressure is measured
3 = Transition to sterilisation phase
when the set temperature is
measured on SelSensor, (set
temperature – Sensor2Dif) is
measured on SelSensor2 and the
associated steam pressure is
measured
4 0, 1, 2, 3 2
33 Sensor2Dif
(Cycle 23)
Permissible deviation between
SelSensor and SelSensor2 when both
are active during control (Heat2SterMod
1 or 3)
4 0.1 – 50.0 °C 0.5 °C
34 SterPressAdd
(Cycle 32)
Additional pressure on the corresponding
steam pressure according to the
sterilisation temperature in the
sterilisation phase
4 0.0 – 100.0 kPa 5.0 kPa
35 LimPrsAdd
(Cycle 15)
Maximum additional pressure:
When this value is reached, the steam
supply from the internal generator stops.
With external steam, HeatGenPrsF is the
limiting factor used to stop the steam
supply (“Steam to chamber” valve is
closed).
4 0.0 – 100.0 kPa 30.0 kPa
36 HeatGenPrsF
(Cycle 10)
Factor for increased steam pressure in
the steam generator.
Ensures that the steam temperature is
adequate.
4 0 – 50% Depends
on program
37 SterGenPrsF
(Cycle 16)
Factor for increased steam pressure in
the steam generator during the
sterilisation phase.
Ensures that the steam temperature is
adequate.
4 0 – 50% Depends
on program
38 CoolPressF
(Cycle 35)
Auxiliary pressure regulated during the
cooling phase according to the set
sterilisation temperature if ExhMode is 3,
4, 5, 6 or 8 and compressed air is
connected.
4 100 – 200% Depends
on program