39 CoolMinPress
(Cycle 38)
Defines whether a minimal amount of
auxiliary pressure should remain in the
sterilisation chamber until the end of the
program.
This function is only possible if ExhMode
is set to 5, 6 or 8.
4 0.0 – 300.0 kPa Depends
on program
40 SaveWtrTemp
(Cycle 61)
Optimises the cooling water usage
during the cooling phase. As long as the
temperature of the draining liquid
measured is higher than that set, the
cooling water valve stays open.
4 10.0 – 110.0 °C 40 °C
41 SaveWtrRate1
(Cycle 62)
If the temperature in the drain is less
than SaveWtrTemp and the product
temperature is higher than 100 °C, the
cooling water valve is pulsed according
to the set parameter.
4 0 – 100% 50%
42
SaveWtrRate2
(Cycle 63)
If the temperature in the drain is less
than SaveWtrTemp and the product
temperature is between 90 and 100 °C,
the cooling water valve is pulsed
according to the set parameter.
4 0 – 100% 40%
43 SaveWtrRate3
(Cycle 64)
If the temperature in the drain is less
than SaveWtrTemp and the product
temperature is less than 90 °C, the
cooling water valve is pulsed according
to the set parameters.
4 0 – 100% 30%
44 CoolTime
(Cycle 36)
Defines the minimum cooling time if the
temperature sinks at more than the
“CoolRateTemp” per minute during
cooling (ExhMode 3 to 6.8).
(reference container broken?)
4 0 – 9999 min. 10 min.
45 CoolRateTemp
(Cycle 65)
Describes the maximum temperature
drop per minute at the control sensor
during the cooling phase if “ExhMode” is
set to 3 – 6 or 8:
If the temperature drops by more than
the set value per minute then the cooling
phase is only ended after the “CoolTime”
has expired, regardless of the measured
temperature.
4 1.0 – 100.0 °C 30.0 °C