8-8 Revision: 2.1
No.
Parameter Description Access
level
Range of values Factory
setting
46 DryMode
(Cycle 56)
Drying mode:
0 = Evacuation of the chamber during
the drying time, the coil is heated at
the same time.
1 = Pulsed evacuation, the coil is heated
at the same time. Evacuate until the
pressure is less than “DryVac”, then
break the vacuum with sterile, filtered
air and restart the evacuation, etc.
2 = Evacuation to the set value “DryVac”,
coil is not heated (e.g. autoclaving
waste).
3 = Drying with compressed air, the coil
is heated at the same time.
4 0, 1, 2, 3 Depends
on program
47 DryVac
(Cycle 67)
Value of the vacuum to be reached in the
drying phase.
4 5.0 – 150.0 kPa Depends
on program
48 BioHazard
(Global 11)
Only for exhaust filters:
The accrued condensation is collected in
the sterilisation chamber and also
sterilised. It is only drained off after the
sterilisation phase if no fault occurs. After
150 cycles, the message “Replace filter”
will appear in the display.
0 = Function inactive
1 = Function active
4 0, 1 Depends
on device
49 VacuumRate
(Cycle 58)
Defines the rate of evacuation.
Used if the chamber is evacuated and
sterilised using filters.
4 0 – 100% 100%
50 VBreak Rate
(Cycle 59)
Defines the rate of breaking the vacuum.
Used if the chamber is evacuated and
sterilised using filters.
4 0 – 100% 100%
51 ExhaustRate
(Cycle 60)
Defines the rate at which steam is vented
after a steam pulse.
Used if the chamber is evacuated and
sterilised using filters.
4 0 – 100% 100%
52 DrnCoolTemp
(Global 14)
Defines the maximum permissible
temperature in the drain.
4 2.0 – 140.0 °C 70 °C
53 CoolDrainOn
(Global 17)
Pulsing interval for the valve to add
cooling water if the temperature set in
“DrnCoolTemp” is exceeded.
4 0 – 100 x 0.1 sec.
80
54 CoolDrainOff
(Global 18)
4 0 – 100 x 0.1 sec.
20
55 SterPrintT
(Cycle 18)
Print interval time for the printer in the
sterilisation phase.
4 0 – 3600 sec. 30 sec.
56 PrintTimer
(Cycle 9)
Print interval time for the printer outside
the sterilisation process.
4 0 – 600 sec. 120 sec.
57 EndCycleBipT
(Global 13)
Defines the length of the audible signal
at the end of the entire process.
4 0 – 9999 sec. 10 sec.
58 HighTempLim
(Cycle 12)
Permissible maximum temperature in the
pressure chamber. If a higher
temperature is measured, the program
will abort.
4 30.0 – 150.0 °C
(optionally up to
152 °C)
145 °C
(optionally
up to
150 °C)