Revision: 2.1 8-9
No.
Parameter Description Access
level
Range of values Factory
setting
59 HeatTimeOut
(Cycle 11)
Maximum time for the heating phase
before the sterilisation temperature is
reached.
4 0 – 9999 min. 300 min.
60 TempMaxErr
(Cycle 24)
The maximum value by which the set
sterilisation temperature may be
exceeded.
4 0.0 – 20.0 °C 3.5 °C
61 TempMinErr
(Cycle 25)
The maximum value by which the set
sterilisation temperature may be fallen
short of.
4 0.0 – 20.0 °C 1.0 °C
62 PressMaxErr
(Cycle 26)
The maximum value by which the steam
pressure belonging to the set sterilisation
temperature may be exceeded.
4 0.0 – 100.0 kPa 50.0 kPa
63 PressMinErr
(Cycle 27)
The maximum value by which the steam
pressure belonging to the set sterilisation
temperature may be fallen short of.
4 0.0 – 20.0 kPa 10.0 kPa
64 TempMaxTime
(Cycle 28)
Time after which an alarm is triggered
when the set maximum temperature is
exceeded.
4 0 – 60 sec. 20 sec.
65 TempMinTime
(Cycle 29)
Time after which an alarm is triggered
when the set minimum temperature is
fallen short of.
4 0 – 60 sec. 20 sec.
66 PressMaxTime
(Cycle 30)
Time after which an alarm is triggered
when the set maximum steam pressure
is exceeded.
4 0 – 60 sec. 20 sec.
67 PressMinTime
(Cycle 31)
Time after which an alarm is triggered
when the set minimum steam pressure is
fallen short of.
4 0 – 60 sec. 20 sec.
68 PulseTimeOut
(Cycle 54)
Maximum time for a pre-vacuum pulse or
steam pulse.
4 0 – 9999 sec. 2400 sec.
69 GenMaxPress
(Global 2)
Permissible maximum pressure in the
steam generator.
4 0 – 700 kPa 400 kPa
70 GenWtrErrT
(Global 4)
Time within which water must be
detected by the top electrode in the
steam generator after the start of water
intake.
4 0 – 200 sec. 120 sec.
71 ErrRingOpenT
(Global 8)
Time within which the RingCloseSw
switch must be activated after the signal
to open or close the door is given.
4 0 – 10 sec. 10 sec.
72 RingPrsAdd
(Global 7)
The value in excess of the atmospheric
pressure (ATMPressure)
within which the cover can be opened.
4 0.0 – 20.0 kPa 4.0 kPa
73 OpenDoorTime
(Global 9)
Time in which the cooling coil is blown
out with compressed air whilst the door is
opening.
4 0 – 200 sec. 30 sec.
74 ATMPressure
(Global 0)
Entry of the atmospheric pressure at the
installation position of the autoclave.
4 50.0 – 104.0 kPa 102.0 kPa