5. INSTALLATION INSTRUCTIONS
5.1. SPECIFICATIONS AND REQUIREMENTS
For correct machine installation, make sure that:
1. The machine is installed on a level surface that is free of vibrations and has sufficient carrying
capacity for the combined weight of the machine and user.
2. The place of installation is free of dust and sand.
3. The place of installation meets the operating temperature and humidity conditions specified in
paragraph 2.5. “Ambient specifications”.
4. The machine is not positioned close to sources of heat, sources of electromagnetic noise
(television sets, electrical motors, antennas, high voltage lines, household appliances, etc...) or
medical equipment.
5. To eliminate any interference with the cardio receiver, no transmitters should be placed less
than 1 meter from the display.
6. There is plenty of free space around each item of equipment and a free space of 2x1 min front of
the machine as shown in the picture:
7. The mains voltage must match the value specified on the machine rating plate.
8. The electrical system must be correctly earthed.
9. The wall outlet used should be reserved for the machine and have a rating of at least 2600 Watt.
10. Do not connect other machines or loads to the same wall outlet.
11. Position the mains lead of the machine where it will not be underfoot.