RUN MED: Service & Maintenance Manual - rev. 1.0
6.3. THE DISPLAY FAILS TO ILLUMINATE
This error occurs when the power supply voltage does not reach the upper assembly.
Does the wall outlet supply the
correct v oltage?
Is the mains lead OK?
Is the voltage at the
output of the machine's
power inlet block correct?
Plug the machine into a
working electrical outlet
Replace the mains lead
Replace the power
inlet block on the
machine
THE DISPLAY
FAILS TO
ILLUMINATE
Is the voltage at the input
to the circuit breakers
correct?
NO
YES
NO
YES
NO
YES
NO
YES
2
A
Replace cable TRM-58
1
Continued on the following page.
Page 6.7