RUN MED: Service & Maintenance Manual - rev. 1.0
Page 6.22
6.6.7. ERROR PE
This error indicate problem on the data of the Eeprom.
Does the error occur
f requently?
YES
Replace the AT driver
board
Probable interf erence. The
machine is OK
NO
Does the machine resume
correct operation af ter being
turned of f f or 1 minute?
YES
Re-program the AT driver
NO
1
Does the machine resume
correct operation af ter being
turned off for 10 minutes?
Programming correct. The
machine is OK
Replace the AT driver
board
NO
YES
PE
2
Follow the procedure step by step to correctly diagnose the problem. Take particular care with the
checks highlighted by circled numbers, which are described in detail below:
(1) Determine the frequency of the errors by counting the number of occurrences and by running
targeted checks. An error is considered frequent if it occurs 2 or 3 times a day.
(2) Re-program the AT driver following the procedure described in paragraph 9.5. “Default
parameters”.