EasyManua.ls Logo

Technogym Run MED - Page 65

Technogym Run MED
150 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
RUN MED: Service & Maintenance Manual - rev. 1.0
Do the pulses correctly reach
the AT driv er board input?
YES
NO
Replace the encoder
4
A
Replace AT driv er
board
Is the 400VAC v oltage
correctly supplied by the AT
driv er board?
YES
Replace the power
supply board
3
NO
Replace the AT driver
board
B
Follow the procedure step by step to correctly diagnose the problem. Take particular care with the
checks highlighted by circled numbers, which are described in detail below:
(1) Place the tester probes across pins 3 and 1 of the connector J6 on the AT driver board. The
measured voltage should be 5 Vdc.
(2) Using a tester, check the voltages on the back of connector XU1 on the AT drive board,
placing the probes on test points 3-4 to measure the 5 Vdc signal (see figure below).
(3) Check that the red LED H5, on the AT driver board is on.
(4) Place the tester probes across pins 2 and 4 of the connector J6 on the AT driver board. The
measured value should be approximately 300-400 mV when the motor is running. If an
oscilloscope is available, it is possible to view the pulses produced by the encoder itself.
Page 6.25

Table of Contents

Related product manuals