Behavior
Use this sequence to trigger the instrument on the nth occurrence of the B event after the A event.
When you select Trig on nth Event, the instrument looks for the trigger conditions specified by the
A event. After the A event arms the trigger system, the instrument looks for a specific number of
B events before triggering and displaying the data.
Trigger on Nth Event (Horizontal Delay On)
To Use
Set up the A Event and B Event from their individual tabs. Then select the A–>B Seq tab and click
Trig on nth Event. Select the number of B events with the Trig Event control. Use the Horiz Delay
control to delay the acquisition relative to the B event. Adjust the Ref Position and B Trig Level
controls as necessary.
Behavior
When you select Trig on nth Event, the instrument looks for the trigger conditions specified by the
A event. After the A event arms the trigger system, the instrument looks for a specific number of
B events before triggering. The instrument then delays the acquisition for the amount of time
specified by the Horiz Delay control. After the delay time has expired, the instrument displays the
data.

Table of Contents

Other manuals for Tektronix TDS5000B Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Tektronix TDS5000B Series and is the answer not in the manual?

Tektronix TDS5000B Series Specifications

General IconGeneral
BrandTektronix
ModelTDS5000B Series
CategoryTest Equipment
LanguageEnglish

Related product manuals