300
Select the color palette
13
Select the Display Setup control window Color tab.
14 Click a Color Palette button in the Fast/Acq WfmDB Palette. In
Fast Acquisitions mode, Temp and Spectral show more detail
than the other color palettes.
For a definition of each color palette, see page 145.
Time Stamping Frames
Use Time Stamps to display the absolute trigger time for a specific frame and the relative time
between triggers of two specified frames. To use FastFrame Time Stamps, do the following steps:
Prerequisites
1
FastFrame mode should be set up as described in the
previous example.
2 Turn on FastFrame
Turn readouts on or off
3
In the Time Stamps controls, click Readouts to toggle time
stamp readouts on or off: Readouts use the following format:
Sel Ch# F xxx DD MMM YYYY
HH:MM:SS.mmm µµµ nnn ppp
Ref Ch# F xxx DD MMM YYYY
HH:MM:SS.mmm µµµ nnn ppp
∆ DD MMM YYYY HH:MM:SS.mmm µµµ nnn
ppp
Where:
Sel and Ref Ch# F xxx are the selected or reference
frame number
DD MMM YYYY is the date (day, month, and year)
HH:MM:SS, is the clock time (hours, minutes, and
seconds)
mmmï€ ÂµÂµÂµ,nnn,ppp is a fraction of a second (to
picoseconds)
Select the reference frame
4
In the Time Stamps controls, click Source and select the
source of the reference frame.
5 In the Time Stamps controls, click Frame and use the
multipurpose knob or keypad to enter the number of the
reference frame. This value sets the starting frame when
measuring the relative time between two frames.