Appendix A: Specifications
A-6
TDS5000 Series User Manual
Table A- 2: Channel input and vertical specifications (Cont.)
Characteristic Description
Lower frequency limit, AC coupled <10Hzfor1M, reduced by a factor of ten when using
a 10X probe; <200 kHz for 50
Calculated rise time, typical DC 50 coupling, bandwidth limit set to Full
TDS5052/TDS5054 SCALE range Rise time
1 mV/div to 1.99 mV/div 2.29 ns
2 mV/div to 4.98 mV/div 1.33 ns
5mV/divto1V/div 800 ps
TDS5104 SCALE range Rise time
1 mV/div to 1.99 mV/div 2.29 ns
2mV/divto1V/div 300 ps
Step response settling errors,
t
i
l
Bandwidth limit set to Full
t
y
pi
c
a
l
SCALE range and step
amplitude
Settling error at time after
step
1 mV/div to 99.5 mV/div,
with 2Vstep
20 ns: 0.5%
100 ns: 0.2%
20 ms: 0.1%
100 mV/div to 1 V/div,
with 20 V step
20 ns: 1.0%
100 ns: 0.5%
20 ms: 0.2%
1.01 V/div to 10 V/div,
with 200 V step
20 ns: 1.0%
100 ns: 0.5%
20 ms: 0.2%

Table of Contents