EasyManua.ls Logo

Terumo CAPIOX EBS Circuit with X coating - Page 115

Default Icon
177 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Various menu settings
2. Tap the channel (P1 or P2) you want to set. (Figure 3-8)
The following figures show the example when [P1] is selected.
<<Setting the on/off status of the pressure measuring
function>>
3. Tap [On] or [Off]. (Figure 3-9)
When turning it off, tap [OK] to complete the setting.
Figure 3-8
<<Setting the threshold value of the Pressure Alarm>>
4. Confirm that [Alarm] of the upper limit is selected.
(Figure 3-10)
Figure 3-9
5. Tap or to input the upper limit value you want to set.
(Figure 3-11)
(red)” indicates the threshold value of the Over Pressure Alarm in
accordance with an input pressure value.
Figure 3-10
6. Tap [Alarm] of the lower limit. (Figure 3-12)
Figure 3-11
Figure 3-12
86

Table of Contents

Related product manuals