Make sure that the underside of the centrifugal
pump is in close contact with the pump receptacle
and that the slide hook is in position (Figure 6).
Make sure that the rim of back side of the
oxygenator and the hook of the exclusive holder
are connected to the slide hook (Figure 7).
Only when the display of the motor rpm indication
shows 0”, the centrifugal pump can be attached or
detached .
6.
Connect the water lines (inner diameter of the tube;
12.7mm (1/2”) or couplers) to the water ports of CAPIOX
EBS circuit (Figure 8-a, 8-b).
7.
Start water circulation through heat exchanger and circulate
for at least 5 minutes. Check for leaks.
Do not use an oxygenator that leaks. Replace it with
another EBS circuit.
8.
Connect the thermistor probe and the temperature monitor
cable (Figure 8-c).
For a monitor, Y.S.I. 400 Series manufactured by
Measurement Specialties Inc. can be used. Use it
with its exclusive cable.
9.
Connect the gas line (inner diameter of the tube; 6.4mm (1/4”))
to the gas port (GAS IN) (Figure 8-d).
Do not block the gas port (GAS OUT).
WARNING
CAUTION
CAUTION
CAUTIONS

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Terumo CAPIOX EBS Circuit with X coating and is the answer not in the manual?

Terumo CAPIOX EBS Circuit with X coating Specifications

General IconGeneral
BrandTerumo
ModelCAPIOX EBS Circuit with X coating
CategoryMedical Equipment
LanguageEnglish

Related product manuals