Make sure that the underside of the centrifugal
pump is in close contact with the pump receptacle
and that the slide hook is in position (Figure 6).
・
Make sure that the rim of back side of the
oxygenator and the hook of the exclusive holder
are connected to the slide hook (Figure 7).
・
Only when the display of the motor rpm indication
shows “0”, the centrifugal pump can be attached or
detached .
6.
Connect the water lines (inner diameter of the tube;
12.7mm (1/2”) or couplers) to the water ports of CAPIOX
EBS circuit (Figure 8-a, 8-b).
7.
Start water circulation through heat exchanger and circulate
for at least 5 minutes. Check for leaks.
Do not use an oxygenator that leaks. Replace it with
another EBS circuit.
8.
Connect the thermistor probe and the temperature monitor
cable (Figure 8-c).
For a monitor, Y.S.I. 400 Series manufactured by
Measurement Specialties Inc. can be used. Use it
with its exclusive cable.
9.
Connect the gas line (inner diameter of the tube; 6.4mm (1/4”))
to the gas port (GAS IN) (Figure 8-d).
Do not block the gas port (GAS OUT).
WARNING
CAUTION
CAUTION
CAUTIONS