10.
After the circuit was assembled, confirm all parts are
connected properly, and the tubes are not folded or kinked.
11.
When using with CAPIOX SP Console SP-101:
Attach the CAPIOX SP Pump Flow Sensor 3/8” to the flow
connector on the arterial line following the Instructions for Use
of CAPIOX SP Pump Console SP-101.
Ultrasonic sensor gel must be fully applied to the
flow connector.
When
a
flow
connector
is
connected
with
a
flow
sensor, attach recommended ultrasonic gel
(AQUASONIC 100 made by Parker) adequately
between the flow sensor and flow connector. Do
not use gel which solvents are included such as
alcohol, thinner and amine, etc. These solvents
may cause damage the sensor and connector.
12.
When using with CAPIOX Centrifugal Pump Controller
SP-200:
Attach the ABD/Flow Sensor to 3/8” tubing, following the
Instruction for Use of CAPIOX Centrifugal Pump Controller
SP-200.
When attaching the ABD/Flow Sensor, apply
enough vaseline to evenly coat the detection
window of the clamp-on ABD/Flow sensor. If the
amount of vaseline applied is insufficient, the Flow
Signal Unstable Alarm may be issued, and air
bubbles may not be detected.
13.
After the circuit was assembled, confirm all parts are
connected properly, and the tubes are not folded or kinked.
CAUTION
CAUTIONS

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Terumo CAPIOX EBS Circuit with X coating and is the answer not in the manual?

Terumo CAPIOX EBS Circuit with X coating Specifications

General IconGeneral
BrandTerumo
ModelCAPIOX EBS Circuit with X coating
CategoryMedical Equipment
LanguageEnglish

Related product manuals