Various menu settings
• Use USB memory drives formatted in the form of FAT or FAT 32.
• Do not connect equipment other than USB memory drives to the USB port.
• When [Cancel] is tapped while exporting log data, the export cancels.
• While exporting log data, the Drive Motor cannot be used.
• While exporting log data, a new log cannot be recorded in the history.
• When a USB memory drive is removed or the power is turned off while exporting, the USB memory drive may be damaged.
• The export status of the log may remain at 0% for several tens of minutes.
Log data
• Log data collected from power on till power off is counted as a single case.
• In the system, the log data of each case is stored separated by measured data such as flow values and the event data such as
alarms/alerts.
• The maximum number of log data stored in the system is 50 cases, 600,000 data points or 120,000 events. When any of these
reaches the upper limit, the oldest log is overwritten with the newest log.
Logs of measured data
• Logs are usually stored each minute. However, logs are stored each second in the following cases:
1) Alarm status
2) When the flow value, pressure value and temperature value are close to the threshold values
3) While operating the motor speed knob
• If the system is stored with the power on, the log data will reach the maximum storage number, and the oldest data will be
deleted and overwritten.
Logs of event data
• Stored when the alarm is issued and cancelled.
98
Notes
Note

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Terumo CAPIOX EBS Circuit with X coating and is the answer not in the manual?

Terumo CAPIOX EBS Circuit with X coating Specifications

General IconGeneral
BrandTerumo
ModelCAPIOX EBS Circuit with X coating
CategoryMedical Equipment
LanguageEnglish

Related product manuals