Table 66 Default Values For Melting Height Operating Parameters
Parameter Scientific Units Input
Sample Size
25 pulses
25
Flag Word
0007 Hex
Log Slope
0.03 dB/LSB
1966
LOG Threshold
0.5 dB
8
CCOR Threshold
25.0 dB
400
Signal Quality Index Threshold
0.5 (dimensionless)
128
SIG Threshold
10.0 dB
160
Calibration Reflectivity
22.0 dBZ
352
Gas Attenuation
0.016 dB/km
1600
Zdr Offset (GDR)
0.0 dB
0
LDR Offset (XDR)
0.0 dB
0
AGC Integration Period
8 pulses
8
Radar Wavelength
5.3 cm.
5300
Dual PRF Filter Stabilization
10 pulses
10
UnCor Refl. Thresh. Control Flag LOG AAAA Hex
Cor Refl. Thresh. Control Flag
LOG and CSR
8888 Hex
Velocity Thresh. Control Flag
SQI and CSR
C0C0 Hex
Width Thresh. Control Flag
SQI and CSR and SIG
C000 Hex
Zdr Refl. Thresh. Control Flag LOG AAAA Hex
AZ/EL Angle Offsets
0°
0000 Hex
Altitude of radar
0 meters MSL
0
8.5 Interface Input/Output Test (IOTEST)
Use IOTEST to test the input and output data busses of the signal processor interface.
When issued, the command causes 16 words to be read from the host controller, after which
the same 16 words are printed.
Typically, the controller supplies a "barber pole" input sequence consisting, for example, of
successive powers of two. If all of the output words are correct, one may conclude that there
are no malfunctioning bits in the interface hardware
RVP900 User Guide M211322EN-J
250