15 14 13 12 11 10 9 8 7 6 5 4 3 2 1 0
---------------------------------------------------------------
| | | | | | | | | | | | | | | |
| TSOUT | Sub Type |Unfold | | 1 1 | 0 0 1 1 |0 | Command
|---|---|----------------|-------|---|-------|-------------------|
When the TSOUT bits select Power Spectrum then, depending on the current major
mode, a further choice may be needed to select one of several spectral view points. The
following table shows the values for the random phase major mode the possible values of
Sub Type.
Table 69 TSOUT Random Phase Major Mode Values
First Trip Value First Trip Description Second Trip Value Second Trip Description
0
Raw First Trip
4
Raw Second Trip
1
Whitened First Trip
5
Whitened Second Trip
2
Cleaned First Trip
6
Cleaned Second Trip
3
Final First Trip
7
Final Second Trip
When the TSOUT bits select Time Series then, a further choice may be needed to select
the time series from the first or second pulse when using the Hybrid Pulse Compression
scheme. For the Random Phase major mode the possible values of Sub Type are:
0
Main pulse time series
1
Second pulse time series (if hybrid pulse)
When time series output is selected the output data consist either of (3xBxN) or (2xBxN)
words, depending on the output format, where B is the number of bins in the current range
mask and N is the number of pulses per ray. Data samples for each bin of pulse #1 are
output
first, followed by those for each bin of pulse #2, and so on up to pulse #N. The data
are output in the same time-order that they were acquired.
In the
floating point format, three words are used for each bin:
15 14 13 12 11 10 9 8 7 6 5 4 3 2 1 0
---------------------------------------------------------------
| | | | | | | | | | | | | | | | |
| Exponent | S | Mantissa | (I)
|------------------|---|----------------------------------------|
RVP900 User Guide M211322EN-J
264