Do not compute a noise sample, but rather, restore the powerup noise defaults.
15 14 13 12 11 10 9 8 7 6 5 4 3 2 1 0
---------------------------------------------------------------
| | | | | | | | | | | | | | | | |
| Starting Range in km (Max 992km) of 32km Sampling Interval | Input 1
|---------------------------------------------------------------|
15 14 13 12 11 10 9 8 7 6 5 4 3 2 1 0
---------------------------------------------------------------
| | | | | | | | | | | | | | | | |
| Internal Trigger Rate (6Mhz/N) to use During Noise Sampling | Input 2
|---------------------------------------------------------------|
The following input words are optional, only if Action=1.
15 14 13 12 11 10 9 8 7 6 5 4 3 2 1 0
---------------------------------------------------------------
| | | | | | | | | | | | | | | | |
|0 0 | (MSB) Log of Measured Noise Level(LSB) | Input 3
|------|--------------------------------------------------------|
The is the same number as GPARM output 6. See 8.8 Initiate Processing (PROC) (page 255)
and 8.10 Get Processor Parameters (GPARM) (page 271).
15 14 13 12 11 10 9 8 7 6 5 4 3 2 1 0
---------------------------------------------------------------
| | | | | | | | | | | | | | | | |
| Noise Level Standard Deviation (in 1/100 of a dB) | Input 4
|---------------------------------------------------------------|
This is the same number as the GPARM output 49.
15 14 13 12 11 10 9 8 7 6 5 4 3 2 1 0
---------------------------------------------------------------
| | | | | | | | | | | | | | | | |
| Ratio of Horizontal/Vertical Noise Power in Hundredths of dB | Input 5
|---------------------------------------------------------------|
This is the same number as the GPARM output 50.
RVP900 User Guide M211322EN-J
254