EasyManua.ls Logo

Vaisala RVP900 - Get Processor Parameters (GPARM)

Vaisala RVP900
484 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
The available operations are:
0
Disable the use of simulated data. RVP900 returns to acquisition and processing of
live data from the A/D converters.
1
Enable processing of simulated data. Subsequent PROC commands use the data
supplied in the next N (N = sample size) LSIMUL with Operation= 2.
The receiver noise and oset levels which are internally maintained by RVP900 are set
to their special simulated values (from the M+ setup menu) by this command. This is
because the measured
osets are not relevant to the simulated data, and must not be
used in the subsequent computations. It is important to issue the SNOISE command
before resuming the acquisition and processing of live radar data.
2 or 3
Load one pulse of data samples beginning with the following 4-word header, and
continuing with an array of items each representing a single instantaneous sample of
(I,Q) data. You may specify one or more bins to be loaded, and RVP900 replicates
these data as necessary in order to fill out the entire count of acquired bins. If the
number of bins is 0, then a zero-valued sample is applied for all channels.
15 14 13 12 11 10 9 8 7 6 5 4 3 2 1 0
---------------------------------------------------------------
| | | | | | | | | | | | | | | | |
| Number of Bins of Simulated Data Which Follow | Input 1
|----------------------------------------------------------------|
15 14 13 12 11 10 9 8 7 6 5 4 3 2 1 0
---------------------------------------------------------------
| | | | | | | | | | | | | | | | |
| Transmit Phase of This Pulse (16-bit Binary Angle) | Input 2
|----------------------------------------------------|-----------|
15 14 13 12 11 10 9 8 7 6 5 4 3 2 1 0
---------------------------------------------------------------
| | | | | | | | | | | | | | | | |
| Transmit Power in Hundredths of dB (Zero dB Nominal) | Input 3
|----------------------------------------|---|-------------------|
15 14 13 12 11 10 9 8 7 6 5 4 3 2 1 0
---------------------------------------------------------------
| | | | | | | | | | | | | | | | |
| Reserved | Input 4
|----------------------------------------------------|-----------|
RVP900 User Guide M211322EN-J
290

Table of Contents