EasyManua.ls Logo

Vertiv Liebert STS2 - Figure 67 Source Setpoints

Vertiv Liebert STS2
160 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Vertiv | Liebert® STS2/PDU
User Manual | Rev. 6 | 10/2017 102
Source Setpoints
The Source 1/2 Setpoints allows you to configure trigger points for each input source (1 or 2) for the Liebert
STS2/PDU via a keypad dialog or HI/LO dialog box. The source numbers do not designate the source as the
preferred or alternate source. The numbers simply identify the input source. See 10.2 - Manual Transfer /
Preferred Source Selection to set the preferred source.
The Setpoints are based on values in relation to the nominal voltage required for the Liebert STS2/PDU. UV
refers to under voltage, meaning the amount of voltage below the nominal voltage. OV refers to over voltage or
the amount of voltage over the nominal voltage.
Fast UV Setting (%) — if the voltage drops below this set percentage of the nominal voltage, the Liebert STS2/PDU
transfers from the current source to the opposite source.
Slow UV Setting (%) — if the voltage remains this set percentage below the nominal voltage for the period of the slow
detection delay, the Liebert STS2/PDU transfers from the current source to the opposite source.
Slow UV detection delay — the number of cycles that the source must remain below the Slow UV Setting before the
Liebert STS2/PDU transfers.
OV Setting (%) — if the voltage exceeds this set percentage of the nominal voltage for the period of the OV detection
delay, the Liebert STS2/PDU transfers from the current source to the opposite source.
OV Detection Delay — the number of cycles that the source must remain above the OV Setting before the Liebert
STS2/PDU transfers sources.
I-PK Xfer Lockout — if current from the source exceeds this threshold, the Liebert STS2/PDU disables source transfers
and has to be reset either manually or automatically. The type of reset is configured under the User Settings.
To configure the setpoints for each source:
1. Select SOURCE 1/2 SETPOINTS from the User Settings dialog box. The Source 1/2 Setpoints dialog is displayed.
Figure 67 Source setpoints
2. Select 1 to configure the settings for Source 1.

Table of Contents

Other manuals for Vertiv Liebert STS2

Related product manuals