Stratus OCT User Manual PN 2660021134133 A
Introduction
1-9
Stratus OCT Hard Drive Partitions
For Stratus OCT systems shipping with software version 5.0 and later, the hard drive is
partitioned into C: and F: drives. The C: drive is reserved for operating system and Stratus
application files, while the F: drive is reserved for data files. Besides the Stratus database,
the F: drive contains the My Documents folder and the computer Desktop. Thus, files you
save to the Desktop can be found in the F: drive.
WARNING: This warning applies to Stratus OCT systems that shipped with software
version 5.0 and later; these systems have the hard drive partitioned into C: and F: drives.
Do not export or save data to the C: drive of a Stratus instrument. The C: drive
is relatively small and can be filled up quickly, which renders the system
unusable. The F: drive is reserved for data and therefore is relatively large. If
you want to export or save data to the Stratus hard drive, either locally or to a
connected Stratus system on the network, select (a location in) the F: drive as
the target.
WARNING: User Changes to Software or Hardware
The Stratus OCT is a medical device. The software and hardware have been
designed in accordance with U.S., European and other international medical
device standards designed to protect clinicians, users and patients from
potential harm caused by mechanical, diagnostic or therapeutic failures.
Unauthorized modification of Stratus OCT software or hardware (including
peripherals) can jeopardize the safety of operators and patients, the
performance of the instrument, and the integrity of patient data.
Unauthorized modification also voids the instrument warranty.
Approved Software
Note: Carl Zeiss Meditec does not provide technical support for the use of third party
software.
Please refer to the Stratus OCT Technical Support section of our website
(
www.meditec.zeiss.com/stratus) for the current list of approved software, including
updates and patches to the Windows® operating system
and security software.
Note Regarding Network Security: Stratus OCT Software Version 7.0 is compatible
with Windows 2000 Service Pack 4, which is primarily a security update to the
Windows operating system. Therefore, if you connect the Stratus OCT to an office
network, we encourage you to install Service Pack 4. Furthermore, please consider
the information in Networking Guidelines. If you connect the Stratus OCT to the
Internet, which is not recommended, see WARNING: Risks of Internet
Connectivity on page A-2.